Cardiac output in normal subjects: negative versus positive pressure ventilation
Abstract
Spontaneous inspiration and Negative Pressure Ventilation (NPV) produce a negative pleural pressure, and this reduction in intrathoracic pressure is transmitted to the right atrium. In contrast, intermittent Positive Pressure Ventilation (PPV) produces inspiratory increases in intrathoracic pressure and therefore right atrial pressure. The comparison of the cardiopulmonary effects between NPV and PPV has been previously reported in physiologic researches carried out both in animal model and in paediatric patients. Our hypothesis was that in healthy adults the application of NPV by iron lung might result in hemodynamic advantages in comparison with PPV. The hemodynamic effects of NPV and PPV, administered through iron lung and mask-ventilation respectively, were studied in 10 normal subjects. Continuous and non-invasive blood pressure and heart rate measurements were recorded by digital oscillometric photoplethysmography both at baseline and during mechanical ventilation. Iron Lung Ventilation (ILV), using intermittent NPV/CNEEP (Continuous Negative Extrathoracic End-expiratory Pressure) in assist control mode, and mask-ventilation by Bi-Level in ST mode were administered at two different settings (NPV: -15/- 4 and -20/-4 cmH2O), (PPV: +15/+4 and +20/+4 cmH2O) respectively. The measurements lasted three minutes per setting without interruption. The mean values of all hemodynamic variables at baseline, during each step of ventilatory treatment and during the recovery phase, showed no statistically significant difference with ILV, on the contrary, during the application of mask-ventilation there was a significant decrease of diastolic, systolic, dicrotic and mean arterial pressure (p = 0.010, 0.002, 0.027, and 0.002 respectively), of cardiac output (p = 0.002), and dP/dtmax (p = 0.037), whereas stroke volume, pulse rate and systemic vascular resistance did not change. Our data show that both spontaneous breathing and mechanical breathing, during ILV, have the same effect on venous return in healthy subjects. In contrast, mask-ventilation produces transient hemodynamic effects due to a reduction in venous return. These results may represent the starting point for further investigation in patients with critical patients undergoing mechanical ventilation.
Introduction
In order to accomplish ventilation, a pressure difference must be developed intermittently across the lungs. Such pressure, called transpulmonary pressure, represents the difference between the pressure in the alveolus (i.e. inside the lung) and the pleural cavity (i.e. outside the lung). With Positive Pressure Ventilation (PPV), the transpulmonary pressure is increased by making the alveolar pressure more positive; in contrast, with Negative Pressure Ventilation (NPV), the transpulmonary pressure is increased by making the pleural pressure more negative.
Both spontaneous and mechanical ventilation induce changes in intrapleural or intrathoracic pressure and lung volume, which can independently affect atrial filling, ventricular emptying, heart rate and myocardial contractility.
Both spontaneous and mechanical ventilation induce changes in intrapleural or intrathoracic pressure and lung volume, which can independently affect atrial filling (preload), ventricular emptying (afterload), heart rate and myocardial contractility (Cardiac Output, CO). Spontaneous inspiration produces a negative pleural pressure and this reduction in intrathoracic pressure is transmitted to the right atrium. In contrast, Intermittent PPV (IPPV) produces inspiratory increases in intrathoracic pressure and therefore right atrial pressure 1.
The effects of PPV on cardiac function, in healthy volunteers receiving “mask” PPV, 2 3 were first reported by Cournand et al. They showed that right ventricular filling was inversely related to intrathoracic pressure and, as this became more positive and the right ventricular preload fell, it produced a detectable fall in CO 2 3.
Although the hemodynamic effects of NPV have not been extensively studied, most clinicians believe that these effects are opposite to those of PPV, more physiological and more likely to maintain a normal CO 4. The exposure of the entire body (except the airway opening) to NPV would result in the same adverse hemodynamic effects seen with PPV 5. These effects occur because intrathoracic pressure actually is raised relative to body surface pressure, reducing the gradient for venous return. Skabursis et al. 6 and Lockhat et al. 7 have shown that this is not the case when the application of NPV is confined to the thorax and upper abdomen by using cuirass or poncho wrap. Unlike tank ventilator, these machines selectively decrease intratoracic pressure, so that the right atrial pressure becomes more negative (compared to the rest of the body), potentially enhancing the gradient for the venous return. In addition, Shekerdemian et al. 8 9 studied the effects of cuirass NPV on cardiac output in healthy children and in children after simple cardiac surgery, demonstrating that cuirass NPV led to a modest improvement in the CO of healthy children while it is able to achieve a greater improvement in post-operative patients.
It has been reported that CO can be detected using a non invasive method: the Pressure Recording Analytical Method (PRAM) 10, which can derive the arterial-pressure-blood flow relationship from the analysis of the arterial pressure profile in the time domain. It has been demonstrated that, in stable patients 10 and in hemodynamically unstable patients without atrial fibrillation 11, PRAM is a valid alternative to pulmonary artery thermodiluition and it is an accurate method for monitoring CO.
The aim of this study was to evaluate the sequences of changes in CO and haemodynamic parameters by PRAM in a group of normal subjects during non invasive mask ventilation (Mask-PPV) and during NPV with an Iron Lung Ventilation (ILV).
Material and methods
Subjects
Ten healthy volunteers from our unit staff with stable hemodynamic condition, without history of previous cardiac and respiratory diseases and free of respiratory symptoms were enrolled in the study. Three subjects were ex-smokers and all of them had gave up smoking at least three years before the starting of the study. All subjects gave their written informed consent to the procedures.
Study protocol
Methods
All subjects underwent spirometric measurements, in a sitting position and at rest from at least 30 minutes, two hours before the start of the study. Arterial blood pressure and heart rate were taken in a supine position immediately before the start of the study.
Continuous and non-invasive blood pressure and heart rate measurements were recorded. Blood pressure measurements were assessed by digital oscillometric photoplethysmography (Finapres, Ohmeda, Englewood, Colorado, USA) 12-17. Arterial blood pressure measurements were taken at the start and at the end of the protocol with a Riva Rocci sphygmomanometer 18.
All subjects were ventilated first by iron lung and then, after 24 hours, by mask ventilation or the other way round, according to a random assignment.
All subjects were ventilated first by iron lung and then, after 24 hours, by mask ventilation or the other way round, according to a random assignment. Randomisation was carried out using sealed envelopes containing random assignment codes; the codes were generated from tables of random numbers by an operator unaware of the aim of the study. The two methods of ventilation were performed at the same time in the afternoon, both with the subjects lying in supine position. Before starting their ventilatory sessions, all subjects were able to familiarise with both type of ventilators and their relative interfaces for 10 minutes. At the end of these preliminary sessions, the subjects were asked to rest in supine position for 30 minutes before starting of the active session. NPV was administered by a microprocessor based iron lung capable of thermistor triggering to perform assist-control NPV (CA 1001, Officine Coppa Biella, Italy) 19 which exposes the entire body (except the upper airways) to subatmospheric pressure (i.e. negative pressure) during inspiration. PPV was administered by a Bi-Level ventilator (Bi-Pap Vision, Respironics Inc.) connected to a face mask (Respironics) with a plateau valve (Respironics).
The ventilatory treatment was carried out over two days, applying for each subject NPV and PPV in the order assigned by the randomization process. ILV was carried out in an assist-control mode, with Intermittent Negative Pressure Ventilation/Continuous Negative Extrathoracic End-Expiratory Pressure (INPV/CNEEP), at two different settings (INPV/CNEEP -15/-4 cmH2O; and -20/-4 cmH2O). Each setting lasted three minutes without interruption. The time of three minutes was found to be sufficient to detect early, by PRAM, the haemodynamic profile of patients with orthostatic intolerance 16. The phases of measurement were: T1N (basal condition), T2N (-15/-4 cmH2O), T3N (-20/-4 cmH2O) and T4N (recovery phase). The low level of EPAP/CNEEP employed due to the fact that the Bilevel vision used had a default of 4 cmH2O as the lowest level of pressure which can be set. Therefore in order to make comparable the experimental condition we applied the same level of continuous negative end expiratory pressure on ILV. Furthermore, the addition of CPNE to NPV was due also to improve the synchrony between the subject and the ventilator as we previously reported 19.
Positive pressure ventilation was carried out by Bi-Level in S/T mode at two different settings (IPAP/EPAP +15/+4 cmH2O, FiO2 21%; IPAP/EPAP +20/+4 cmH2O, FiO2 21%) each setting lasted three minutes without interruption during the experimental condition. The phases of measurements were: T1P (basal condition), T2P (+15/+4 cmH2O), T3P (+20/+4 cmH2O) and T4P (recovery phase).
Systemic arterial blood pressure, pulse rate and CO during each session of ILV, Mask-PPV and during spontaneous breathing were analysed and recorded by PRAM.
PRAM method
The PRAM method 10 17 differently from other Pulse Counter Methods (PCMs), which need external calibration and/or pre-estimated parameters from other subjects 20-25, can derive cardiac flow and arterial impedance (Z(t)) directly from the analysis of the pressure wave. The Stroke Volume (SV) calculation (based upon an objective analytic description of the profile of the entire cardiac cycle, systolic, diastolic and of pressure over time) reflects the effects of physical forces which act on the whole cardiovascular system, thus determining the relationship between arterial blood pressure morphology profile and blood flow. The PRAM system uses the pressure signal obtained both by invasive (aorta, pulmonary, femoral and radial arteries) and non invasive methods 10 16; in this last case the pressure signal is derived from the arterioles of a finger of the hand by an oscillometric-plethysmographic cuff 10.
In detecting the SV, the PRAM system takes into account both the pulsate and the continuous blood flow contribution without the need of any external calibration or pre-estimated parameters from other subjects 10 17. The pulsate component of the flow is given by the impedance Z(t), mainly linked to the ventricular contractility and to the aorta compliance, whereas the continuous component of the flow is mainly linked to the ventricular compliance interaction with compliance resistances of the arterial system.
The basic principle of the PRAM algorithm is the ability to detect the interactions between cardiac and circulatory systems, and therefore to calculate a new parameter that describes the work expenditure of the cardiovascular system strictly correlated to SV; this is called the Cardiac Cycle Efficiency (CCE) 17 26.
In order to verify that the pressure measurements obtained by the Finapres photoplethysmograph were not influenced by the subatmospheric pressure generated inside the iron lung, we compared these measurements with those obtained with the Riva Rocci sphygmomanometer before, during and at the end of each session.
Data processing
Analog noninvasive pressure signals acquired at a sample rate of 1,000 Hz were digitized with an analogic-digital multifunction card (DAQ Card-700, National Instruments Corp., Austin, TX) working on the tension signals with 12 bit from -2.5 to 2.5 volt. All the signals were recorded on a personal computer (Acer, TravelMate 6,000). Data from pressure signals were evaluated beat to beat in real time (Figure 1).
From the pressure signal, the following parameters were measured and estimated by computerized PRAM algorithm:
- Pulse Rate, PR (bpm);
- systolic, diastolic, dicrotic and mean arterial pressure (mmHg);
- Stroke Volume, SV (mL);
- Cardiac Output, CO (L/min);
- dP/dtmax (mmHg/msec);
- Systemic Vascular Resistance, SVR (dines*sec/cm5).
Statistical analysis
The normal data distribution was assessed by means of Kolmogorov-Smirnov one-sample test. Data were expressed as means ± SD and comparisons were performed by Student’s t-test for paired data (negative vs. positive pressures sessions) and by analysis of variance (ANOVA) and post-test (Bonferroni) (within each experimental session) using PASW statistical package 17.0 (IBM-SPSS, Chicago, IL-USA). p < 0.05 was considered to be statistically significant.
Results
Anthropometric, spirometric and cardiac data are reported in Table I. Six subjects were males and four females with a mean age of 45 ± 3 years (range from 36 to 45). Three subjects had a history of smoking, and all had stopped at least 3 years before the enrolment in the study. In Table II are reported the mean values (± SD) of the hemodynamic variables at basal conditions during spontaneous breathing (T1) and during each phase of the ventilatory treatment (T2, T3) with ILV and Mask-PPV.
During ILV there was no statistical significant difference in all the hemodynamic variables studied from T1N, trough the steps T2N, T3N and T4N (recovery from the end of the session).
During ILV there was no statistical significant difference in all the hemodynamic variables studied.
Opposite to basal condition (T1P), during the application of Mask-PPV (T2P, T3P) and recovery phase (T4P) there was a significant decrease of diastolic, systolic, dicrotic, mean arterial pressure (p = 0.010, 0.002, 0.027, 0.002), CO (p = 0.002) and dP/dtmax (p = 0.037), whereas no modification was found for SV, PR (p = 0.728) and only a slight decrease was detected for SVR (p = 0.06).
No statistical differences were detected in all variables when the positive pressure was increased from 15 to 20 cmH2O. The beat to beat analysis of pressure wave morphology detected by PRAM system was not influenced by the subatmospheric pressures generated in the IL, as showed in Figure 2.
During ILV, blood pressure values resulted similar to those obtained before and after each session. On the contrary, during Mask-PPV all variables except PR showed a progressive decline during the test; this trend remained stable during the recovery session.
In order to evaluate the differences between the mean values of the T2 sessions (T2N vs T2P) and the T3 sessions (T3N vs T3P) during ILV and Mask-PPV, the Bonferroni post hoc test was used. As reported in Table II, only PR did not show a significant between the values obtained using ILV and Mask-PPV in both T2 and T3 phases. During the T2 sessions diastolic, systolic, dicrotic, mean arterial pressure, SV, CO and dP/dtmax showed lower values when the subjects underwent Mask-PPV compared to those obtained with ILV (p = 0.003; 0.001; 0.010; 0.001, 0.001, 0.001, and 0.002 respectively), whereas no modification was detected for PR and SVR.
During the application of Mask-PPV and recovery phase there was a significant decrease of diastolic, systolic, dicrotic, mean arterial pressure, CO and dP/dtmax, whereas no modification was found for SV, PR and only a slight decrease was detected for SVR.
During T3 sessions, the values of diastolic, systolic, dicrotic, mean arterial pressure, CO and dP/dtmax were lower when the subjects underwent Mask-PPV in comparison with those obtained with ILV (p = 0.010; 0.001; 0.023; 0.003, 0.001, and 0.030 respectively), whereas no modification was detected for SV, PR, and SVR (p = 0.056).
Discussion
The results of our study show that in healthy subjects there was a different hemodynamic response to the application of a positive pressure, administered at the airway opening trough a facial mask, or to an equivalent negative pressure generated inside a body ventilator by an iron lung. Given that the hemodynamic variables obtained in our study were derived by the analysis of arterial pressure wave, we took particular attention in constantly comparing the values generated by the Finapres photoplethysmograph system with those measured by the Riva Rocci sphygmomanometer.
In healthy subjects there was a different hemodynamic response to the application of a positive pressure or to an equivalent negative pressure generated inside a body ventilator by an iron lung.
No significant variations of arterial blood pressure were detected during ILV both in phase T2N (-15 cmH2O) and T3N (-20 cmH2O) in comparison with basal values, whereas a statistically significant decrease of arterial pressure was found during Mask-PPV both in phase T2P (+15 cmH2O) and T3P (+20 cmH2O) in comparison with baseline values. Furthermore, after the application of PPV, we found that the subjects were not able to return to their basal hemodynamic values of five minutes after the interruption of the ventilatory session. This was not the case when ILV was applied. What does it mean from an hemodynamic point of view?
Cournand et al. 3 showed that in normal subjects the increase in intrathoracic pressure produced by the application of facial mask delivering Intermittent Positive Pressure Breathing (IPPB), produced a decrease in right ventricular filling and therefore in cardiac output. The decrease in cardiac output (14.5% and 16.5%) was proportionally correlated to the increase in mean/peak mask pressures (7/14 and 10.6/16.7 mmHg respectively) applied in two separate experimental condition. In a subset of subjects who underwent a 70-minute period of continuous IPPB, the decrease in cardiac output was greater during the first 10-minute period (23%) as compared to a 17% decrease present after the 40 and 70-minute periods of IPPB. In our normal subjects the application of a pressure of 15 and 20 cmH2O resulted in a decrease in cardiac output of 14.3% and 9% from baseline values. This trend is partially in agreement with the results reported by Cournand 3. Differently by Cournand we found a decrease in the diastolic, systolic and mean arterial pressure measurements during the application of Mask Ventilation. This may be due to a different duration of the ventilatory sessions, 10 minutes in the Cournand protocol versus 3 minutes in our protocol. However, using a shorter period of ventilation we were able to detect the initial perturbation effect on the vascular tone which can be explained by the decrease in diastolic pressure. On the contrary, the lack of significant changes in blood pressure in Cournand study 3 could be justified by the longer duration of the ventilator session, which does not allow the observation of the transient effect of cardiovascular perturbation which appears during the first few minutes. In fact, what the Authors detected could have been not the transient effect on vascular tone, but a new compensated equilibrium of the vascular system.
In our study, the observed decrease in dicrotic pressure during Mask-PPV may mean that in healthy subjects the adaptation of the cardiovascular system to the decreased venous return consists in a variation of both the time of closure and the form of wave pressure of the aortic valve. This influences the CO without any significant change in PR and, consequently, in SV whereas dP/dtmax decreased significantly (p = 0.037), indicating a potential mechanical stress due to PPV on the venous return.
During ILV no significant changes of the studied variables were observed at the two different setting. These results may be explained by the fact that exposing the whole body, with the exception of the head, to subatmospheric pressure inside the iron lung, the intrathoracic pressure actually generated is raised relative to the body surface pressure, thus reducing the gradient for venous return and the correlated variables. Lockhat et al. 7 showed that the cardiac output of normal dogs receiving thoracoabdominal negative pressure ventilation was significantly improved by augmentation of the systemic venous return, but this phenomenon was not present during the whole body negative pressure ventilation.
In our study, we intentionally applied negative and positive pressure for 3 minutes 16 and we thereafter increased the pressure without interruption, in order to verify the response of the physiologic system to the variation of transthoracic pressure and the type of compensation adopted by the cardiovascular system. Our data clearly show that only with the application of PPV the maximal response, in terms of reduction of cardiovascular variables (diastolic, systolic, dicrotic, mean arterial pressure, SV, CO, PR, dP/dtmax, SVR), was obtained in the first three minutes of ventilation, and that increasing the level of pressure there was no further modification of the variables detected. This may be linked to the compensatory mechanism that is put in place in healthy subjects.
Only with the application of PPV the maximal response, in terms of reduction of cardiovascular variables was obtained in the first three minutes of ventilation.
This is the first study which reports a formal comparison on the hemodynamic effects of the application of non invasive PPV and NPV in healthy subjects. Our data confirm the previous physiologic researches carried out both in animal model 7, on the comparison of PPV and NPV, and in healthy subjects 3 during PPV. Our data show that both spontaneous breathing and mechanical breathing during ILV induce parallel changes in intrapleural or intrathoracic pressure which are transmitted to the right atrium with the same effect on venous return. In contrast, Mask-PPV produces an inspiratory increase in intrathoracic pressure with important transient hemodynamic effects due to a reduction in venous return. These data however must be considered keeping in mind the following limitations of the study: the shorter duration of the application of positive pressures, and particularly the lack of a recovery time between their phases of administration, did not allow us to fully replicate the findings of Cournand. The application of the PRAM system instead of the thermodiluition technique, which represent the gold standard for hemodynamic evaluation; however, the PRAM system has been proven to be comparable with the thermodilution technique in animal models 27, in stable patients 8, septic patients 17 and also in hemodynamically unstable patients without atrial fibrillation 11.
Conclusions
Our data show that both spontaneous breathing and mechanical breathing, during ILV, have the same effect on venous return in healthy subjects. In contrast, mask-ventilation produces transient hemodynamic effects due to a reduction in venous return.
Both spontaneous breathing and mechanical breathing, during ILV, have the same effect on venous return in healthy subjects. In contrast, mask-ventilation produces transient hemodynamic effects due to a reduction in venous return.
These results may represent the starting point for further investigation in critical patients undergoing mechanical ventilation.
Figure e tabelle
Variable | Mean (SD) |
---|---|
• Gender, M/F | 6/4 |
• Age, years | 45(3) |
• Height, (cm) | 173(8) |
• Weight, (kg) | 73(12) |
• FEV1, L; % pred | 3.16 (0.39); 103 (11.8) |
• FVC, L; % pred | 3.82 (0.33); 107 (12.4) |
• FEV1/VC % | 82.5 (4.79) |
• BSA, m2 | 1.86 (0.19) |
• BP SYS, mmHg | 121(13) |
• BP DIA, mmHg | 61(10) |
• HR b/min | 72 (2.5) |
Diastolic PressuremmHg | Systolic PressuremmHg | Dicrotic PressuremmHg | |||||||
---|---|---|---|---|---|---|---|---|---|
NPV | PPV | NPV | PPV | NPV | PPV | ||||
Mean (SD) | p* | Mean (SD) | p* | Mean (SD) | p* | ||||
Stroke Volumeml | Cardiac Outputl/min | Pulse Ratebeats/min | |||||||
NPV | PPV | NPV | PPV | NPV | PPV | ||||
Mean (SD) | p* | Mean (SD) | p* | Mean (SD) | p* | ||||
Mean Arterial PressuremmHg | dP/dtmaxmmHg/msec | Systemic Vascular Resistancedines*sec/cm5 | |||||||
NPV | PPV | NPV | PPV | NPV | PPV | ||||
Mean (SD) | p* | Mean (SD) | p* | Mean (SD) | p* | ||||
T1 | 61.4 (10.8) | 62.3 (9.4) | 0.369 | 120.1 (13.7) | 121.0 (12.1) | 0.642 | 76.8 (12.2) | 76.2 (10.5) | 0.378 |
T2 | 63.8 (7.9) | 50.5 (8.4) | 0.003 | 124.5 (11.8) | 102.2 (11.6) | 0.001 | 79.4 (11.0) | 66.5 (7.6) | 0.010 |
T3 | 63.4 (8.2) | 49.9 (9.1) | 0.010 | 121.0 (13.3) | 100.9 (10.9) | 0.001 | 78.7 (11.1) | 65.5 (8.0) | 0.023 |
T4 | 60.4 (7.0) | 54.4 (7.5) | 0.025 | 114.8 (12.1) | 105.0 (11.4) | 0.004 | 75.9 (8.5) | 69.1 (6.2) | 0.025 |
T1-T4p | 0.782 | 0.010 | 0.319 | 0.002 | 0.787 | 0.027 | |||
T1 | 78.7 (15.2) | 80.5 (14.4) | 0.162 | 5.7 (0.49) | 5.6 (0.42) | 0.194 | 72.9 (10.5) | 73.5 (10.1) | 0.200 |
T2 | 82.2 (18.2) | 73.3 (18.0) | 0.001 | 5.8 (0.69) | 4.9 (0.66) | 0.001 | 72.2 (11.1) | 68.8 (11.6) | 0.174 |
T3 | 82.3 (17.9) | 77.3 (18.4) | 0.063 | 5.7 (0.63) | 5.2 (0.65) | 0.001 | 71.5 (11.0) | 68.7 (11.9) | 0.226 |
T4 | 83.2 (11.4) | 79.6 (15.1) | 0.980 | 5.4 (0.62) | 5.4 (0.55) | 0.655 | 70.1 (8.2) | 70.0 (8.2) | 0.713 |
T1-T4p | 0.956 | 0.769 | 0.617 | 0.002 | 0.903 | 0.728 | |||
T1 | 81.2 (11.1) | 82.0 (10.0) | 0.532 | 0.95 (0.14) | 0.93 (0.09) | 0.291 | 1079(158) | 1073(133) | 0.665 |
T2 | 84.0 (8.3) | 67.6 (7.9) | 0.001 | 0.99 (0.14) | 0.77 (0.20) | 0.002 | 1095(137) | 1030(148) | 0.281 |
T3 | 82.7 (9.3) | 66.9 (8.6) | 0.003 | 0.93 (0.17) | 0.79 (0.18) | 0.030 | 1082(138) | 954(128) | 0.056 |
T4 | 78.3 (7.7) | 71.2 (8.1) | 0.009 | 0.86 (0.15) | 0.78 (0.15) | 0.035 | 1074(108) | 976(109) | 0.021 |
T1-T4p | 0.545 | 0.002 | 0.351 | 0.037 | 0.998 | 0.060 |
Riferimenti bibliografici
- Shekerdemian L, Desmond B. Cardiovascular effects of mechanical ventilation. Arch Dis Child. 1999; 80:475-80.
- Motley HL, Cournand A, Werko L. Intermittent positive pressure breathing. A means of administering artificial respiration in man. JAMA. 1948; 137:370-82.
- Cournand A, Motley HL, Werko L, Richards DW. Physiological studies of the effects of intermittent positive pressure breathing on cardiac output in man. Am J Physiol. 1948; 152:162-74.
- Corrado A, Gorini M. Principles and practice of mechanical ventilation. McGraw Hill: New York; 2013.
- Maloney JV, Whittenberg JL. Clinical implication of pressures used in the body respirator. Am J Med Sci. 1951; 221:425-30.
- Skabursis M, Helal R, Zidulka A. Hemodynamic effects of external continuos negative pressure ventilation compared with those of continous positive pressure ventilation in dogs with acute lung injury. Am Rev Respir Dis. 1987; 136:886-91.
- Lockhat D, Langleben D, Zidulka A. Hemodynamic effects of external continuos negative pressure ventilation compared with those of continuous positive pressure ventilation in dogs wuth acute lung injury. Am Rev Respir Dis. 1992; 146:677-80.
- Shekerdemian LS, Bush A, Shore DF. Cardiopulmonary interaction after Fontan operations: augmentation of cardiac output using negative pressure ventilation. Circulation. 1997; 96:3934-42.
- Shekerdemian LS, Bush A, Shore DF. Cardiorespiratory responses to negative pressure negative ventilation after tetralogy of Fallot repair: a hemodynamic tool for patients with a low-output state. J Am Coll Cardiol. 1999; 33:549-55.
- Romano SM, Pistolesi M. Assessment of cardiac output from systemic arterial pressure in humans. Crit Care Med. 2002; 30:1834-41.
- Zangrillo A, Maj G, Monaco F. Cardiac index validation using the pressure recording analytic method in unstable patients. J Cardiothorac Vasc Anesth. 2010; 24:265-9.
- Peñaz J. Photoelectric measurement of blood pressure, volume and flow in the finger. Digest of the International Conference on Medicine and Biological Engineering. 1973;104.
- Parati G, Casadei R, Groppelli A. Comparison of finger and intra-arterial blood pressure monitoring at rest and during laboratory testing. Hypertension. 1989; 13:647-55.
- Imholz BP, Settels JJ, Van der Meiracker AH. Non-invasive continuous finger blood pressure measurement during orthostatic stress compared to intra-arterial pressure. Cardiovasc Res. 1990; 24:214-21.
- Wesseling KH, De Wit B, van der Hoeven GMA. Physical, calibrating finger vascular physiology for Finapres. Homeostasis. 1995; 36:67-82.
- Romano SM, Lazzeri C, Chiostri M. Beat-to-beat analysis of pressure wave morphology for pre-symptomatic detection of orthostatic intolerance during head-up tilt. J Am Coll Cardiol. 2004; 44:1891-7.
- Morelli A, Singer M, Ranieri VM. Heart rate reduction with esmolol is associated with improved arterial elastance in patients with septic shock. A prospective observational study. Intensive Care Med. 2016; 42:1528-34.
- Romano SM, Chiostri C, Olivotto J. Shortened head-up tilting test guided by systolic pressure reductions in neurocardiogenic syncope. Circulation. 2002; 106:e69.
- Gorini M, Villella G, Ginanni R. Effect of assist negative pressure ventilation by microprocessor based iron lung on breathing effort. Thorax. 2002; 57:258-62.
- Alderman EL, Branzi A, Sanders W. Evaluation of the pulse-contour method of determining stroke volume in man. Circulation. 1972; 46:546-58.
- Wesseling KH, Jansen JRC, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear, three-element model. J Appl Physiol. 1993; 74:2566-73.
- Langewouters GJ, Wesseling KH, Goedhard WJA. The static elastic properties of 45 human thoracic and 20 abdominal aortas in vitro and the parameters of a new model. J Biomech. 1984; 17:425-35.
- Stock WJ, Baisch F, Hillebrecht A. Noninvasive cardiac output measurement by arterial pulse analysis compared to inert gas rebreathing. J Appl Physiol. 1993; 74:2687-93.
- Houtman S, Oeseburg B, Hopman MTE. Non-invasive cardiac output assessment during moderate exercise: pulse contour compared with CO2 rebreathing. Clin Physiol. 1999; 19:230-7.
- Remmen JJ, Aengevaeren WR, Verheugt FW. Finapres arterial pulse wave analysis with Modelflow is not a reliable non-invasive method for assessment of cardiac output. Clin Sci. 2002; 103:143-9.
- Romano SM. Cardiac cycle efficiency: a new parameter able to fully evaluate the dynamic interplay of the cardiovascular system. Int J Cardiol. 2012; 155:326-7.
- Romagnoli S, Romano SM, Bevilacqua S. Cardiac output by arterial pulse contour: reliability under hemodynamic derangements. Interact Cardiovasc Thorac Surg. 2009; 8:642-6.